Title:Hallikainen Instruments Harkins-Bramson Synchronous Electrocardiac Massage Machine
Description:-None-
Keywords:-None-
Meta data:-None-
Body:THE IIABKINS-BBAMSON SYNCHRONOUS ELECTROCARDIAC MASSAGE A Cardiac Remrcitator and Assistor Model 1318 MACHINE

--

The Harkins-Bramson machine, utilizing the Johns Hopkins method of closed-chest cardiac massage, delicotely controls the circulation of patients in cardiac arrest and extends the use of the same principle to support the foil- ing but still beating heart. Thus, for the first time, there is an available means of directly supporting the circulation of patients in heart failure.

EXPLANATION OF PRINCIPLE Kouwenhoven, Jude and Knickerbocker hove demonstrated that pressure over the lower sternum during cardiac arrest compresses the heart against the vertebral column ond forces blood out of the ventricles of the heart. These workers found that o repetitive or pulsating force correctly applied over the sternum produced a sufficient car- diac output to maintain life during cardiac arrest. This method of cardiac massage has become the preferred treat- ment for the emergency of cardiac arrest, for several reasons. It is more quickly applied than open chest massage; does not require surgery; does not lead to infection; con be used by non-professional personnel; and produces better circulation than open-chest cardiac massage.

CARDIAC RESUSCITATOR In the past, external cardiac mossoge has been applied only by hand. When compared to proper mechanical application however, the manual method is distinctly inferior. Application by hand tends to be uneven; too little force will lead to poor circulation and ieopordize success, whereas, too exuberant on application may lead to un- necessary trauma to the thoracic cage or abdominal organs. External cardiac massage by hand is exhausting, fre- quently requiring the services of several operators. Mechanical application is more convenient and more opt to result in o successful outcome. The force applied con be delicately controlled; the minimum pulsating force needed to produce satisfactory cardiac output being consistently applied for whatever period is necessary. An optimum regime, once established, con be continuously maintained. This assures satisfactory circulation with minimum haz- ard of rib fracture or other troumo. Only one operator is needed and he is free to core for other aspects of emerg- ency core, such os attention to airway problems, application of electrical defibrillation, injection of drugs, etc.

CARDIAC ASSISTOR When closed-chest cardiac massage is used for patients in cardiac arrest, whether in asystole or ventricular tibril- lotion, the timing of the applied sternal forces in relation to the heart's activity is unimportant. However, os soon os the heart resumes coordinated activity, randomly applied massage will intermittently interfere with cardiac filling and thereby decrease cardiac output. By exact synchronization of the rhythm of external mossoge with the ventri- cular contractions, the Harkins-Bramson machine is able to increase cardiac output while the heart is actively beot- ing. The instrument is thus o cardiac assistor capable of increasing cardiac output by external means, and is use- ful in the treatment of o variety of conditions of circulatory failure. -1

DESCRIPTION OF THE INSTRUMENT There ore two components, the cardiac massage unit [Fig. 2) and the pulse generator (Fig. 3). The cardiac massage unit is placed astride the thorax of the patient with the massaging pod in contact with the lower ster- num. The posterior brace transmits the necessary reaction forces to the vertebrae directly below the pad. Lateral movement of the patient is prevented by two adjustable wedges, while a divided mattress makes the position comfortable for the subject. A source of compressed gos (wall oxygen, or o cylinder of air, nitrogen, CO*, or 0,) provides the energy that drives the massaging piston pad. The compressed gas is led through o pressure reducing valve into o reservoir at whose outlet o three-way solenoid valve is fitted. When energized, this valve connects the reservoir to the cylinder of the cardiac mossoge unit, and results in a downward force applied through the pod to the sternum. When not energized, the cylinder is exhausted to atmosphere via o second reservoir which serves os o silencer. The force applied to the sternum by the massaging pod con be delicately and continuously controlled by the operator between 0 and 90 Ibs. A circular watertlled bladder fitted to the massaging pad prevents local bruis- ing by applying the necessary force in the form of fluid pressure, uniformly distributed. When the instrument is used os o cardiac assistor, the pulsating force is triggered by the "R" wove of the ECG through on electronic syn- chronizer in such a way that the impressed sternal force is applied to the heart at precisely the instant of ventricular contraction. The duration of imposed systole, which is to soy the time interval during which each mechanical stroke compresses the heart, is also controlled by the operator. This is an essential adjustment, since too short on interval allows insufficient time for ejection of blood from the ventricles, and if too long, it interferes with the re- filling of the heart. When used for the treatment of cardiac arrest (when there is no "R" wove), coordination with the heart is neither possible nor necessary, and the frequency of the pulsating force is then controlled by the opero- tar. The duration of imposed systole is then o preset but adiustable fraction of each cycle. 1

Figure 2

An additional feature of the instrument is a device which automatically switches the timing control from syn- chronization with the KG to CI pre-set frequency, or vice versa, as the patient moves into, or out of, cardiac arrest. This automatic feature ccln be switched on or off by the operator; if it is "on", and if the patient moves into cardiac arrest, the necessa~ change to o pre-set pulse rate is electronically delayed [about 3-4 seconds) sufficiently to dis- tinguish the arrest from CI mere irregularity of pulse. This delay is also adjustable.

Figure 3 APPLICATION a) Resuscilation: Provided that the cardiac arrest does not yield to manual massage immediately, the Harkins-Bramson machine should be applied. The instrument is especially de- vised for quick application. Inasmuch OS manual massage can be maintained while the instrument is adjusted to the patient, no time need be lost in switching from manual to mechanical support. The instrument is then adjusted to provide the op- timal massaging rate and force as determined by peripheral arterial pulsations or blood pressure measurement. With op- timal circulation maintained by the machine, attention can then be directed to other aspects of emergency core. When heart action returns (should the heart go intofibrillation, an external defibrillator should be applied), the rate previously set by the attending physician will immediately be overridden by the patient's own heart rate and massage will coordinate to provide support to the heart not only while arrested but also during the critical moments of recovery following arrest.

Following successful resuscitation, whether accomplished by hand or with the aid of the machine, it is recom- mended that the machine remain connected to the patient on o stand-by basis and that the electrocardiogram be monitored continuously for several hours so os to provide the best safeguard in the event of subsequent heart fail- ure or arrest. b) Support of the Failing Heart; When called upon os an assistor for the beating heart, the instrument will be used on on elective basis and leisurely application is possible. After placing the cardiac massage unit in the car- rect position around the patient, the ECG signals ore fed into the pulse generator and the mossoge commenced in synchrony with the beating heart. The force of massage is gradually increased until cardiac output is sufficiently improved. The control of "duration of systole" is then adjusted so a to yield maximum cardiac output from that force. Thereafter, the force of the massaging pad con be readjusted to the minimum yielding adequate circulatory support.

Summary of Uses The Harkins-Bramson machine is designed for use in two large oreos of clinical medicine-resuscitation of the arrested heart, and support of the circulation during heart failure. a) Cardiac Arresl It is now well established that external massage is the safest means of cardiac resuscitation. In the event of the catastrophe of cardiac arrest, external massage should be applied manually without delay. If spontaneous heart action returns immediately, manual massage should be discontinued because randomly timed external mas- soge will then interfere with cardiac filling. If the heart does not respond promptly to manual massage, the patient should be placed on the massage unit immediately. The proper force and frequency for massage con be established quickly and continued with uniformity for os long os necessary. -_ b) Heart Failure The some principle that permits blood tlow to be initiated in cardiac arrest con be applied to increase cardiac output in the actively beating heart, provided that the mechanical force is applied simultaneously with each ven- tricular systole. The Harkins-Bramson machine accomplishes this by precise electronic synchronization with the electrical excitation wove of the ventricles ("R" wove of the electrocardiograph). External support is provided to each myocardial contraction, thereby increasing stroke volume and cardiac output. By virtue of this feature, the in- strument is the first and only cardiac ossistor currently available. It will support the heart in any case of heart failure due to an abnormality of the heart itself. Examples of its numerous applications in this area of clinical medicine include the treatment of the heart failure of myocardiol in- forction, Stokes-Adams disease, and myocarditis, and post-operative suppoti of the heart after cardiac surgery. Heart failure due to peripheral causes, such OS hemorrhage or septicemia, is not amenable to treatment by o car- disc assistor. It should be pointed out that the treatment of cardiogenic heart failures by the Harkins-Bramson machine repre- sents o mode of therapy that has not previously been available. Although, in some instances, it will replace an- other form of treatment, more often it will offer hope in o situation for which no other treatment is known.

-,